Coulometric Karl Fischer titrators
  • Name:Coulometric Karl Fischer titrators
  • Model:ZYSCKF105
  • Catalogue :Oil Testers
Inquiry

I.              Instrument description

The ZYSCKF105 Coulometric Karl Fischer titrators is an used for Coulometric water determination according to Karl Fischer. The instrument is designed for usage as a titrator in analytical laboratories.Its main application field is volumetric titration. it is the standard of GB/T7600 production.

 

The present instrument is suitable for processing chemicals and flammable samples. The usage of  this titrators therefore requires that the user has basic knowledge and experience in the handling of toxic and caustic substances. Knowledge with respect to the application of the fire prevention measures prescribed for laboratories is also mandatory.

 

II.            Safety instructions

Warning: This instrument may only be operated in accordance with the specifications in this documentation.


Warning: An incorrect mains voltage can damage the instrument.Only operate this instrument with a mains voltage specified for it (see rearpanel of the instrument).


Warning:Never open the housing of the instrument. The instrument could be demaged by this. There is also a risk of serious injury if live components are touched.There are no parts inside the housing which can be serviced or replaced by the user.

 

Technical index:

Name

Data

Endpoint indication

Voltametric, AC indication

Iodine production

Pulse with variable current strength and length

Current at the generator electrode: 100, 200, 400 mA

Titration speed max

2.4 mg H2O/min

 

Determination range

3 ug to 150 mg H2O

Resolution

0.1 ug H2O

Reproducibility Sample

Reagent manufacturer's standard.

With 10 ug...1000 ug H2O: ± 2 ug

With >1000 ug H2O: 0.2% or better

Drift compensation

automatic

Display Graphical LCD

240 x 128 dots, LED back-lit

Printer

Built-in thermal printer, Paper width 57 mm

Memory

Data storage 500 results

Stirrer control

coordinated with the titration process

RS232 interface

computer connection

Ambient temperature

 

 

Nom. operation range  5 ...40 °C

Storage – 20 ...60 °C

Transport – 40 ...60 °C

 

Mains connection

Voltage 100...240 V ± 10 %

Frequency 50 ... 60 Hz

Power consumption max. 30 W

Material of housing

Steel

Material of display

Glass(LCD)

Dimensions

L*W*H: 320mm*270mm*90mm

Weight

8.6 kg